2 info messages, Info messages -2 – Xylem IQ SensorNet ViSolid 700 IQ (SW) User Manual

Page 50

background image

Indexes

ViSolid

®

700 IQ (SW)

8 - 2

ba76040e02

07/2013

8.1.2

Info messages

EC2342

User calibration error,
check TSS/SiO2 pairs of variates
* All TSS values within measuring range?
(see operating manual)
* At least one value pair entered?
* All TSS and SiO2 values entered?
* All TSS/SiO2 pairs in descending order?
* Pair 1 = highest TSS and SiO2 value?

EI1342

Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Defective components, replace components

EI2342

Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components, replace components

ES1342

Component hardware defective
* Contact service

ESD342

SensCheck: Measurement interfered
* Submerse sensor in sample
* Select bubble-free spot for measurement
* Remove any foreign matter from sensor
* Avoid influence of large foreign matter
* Clean sensor
* Increase signal average time

ESE342

SensCheck: Ultrasound cleaning system has failed
* Return sensor for repair

Message code

Message text

Message code

Message text

IA1342

Ultrasound cleaning system switched on
* Check sensor visually
* Clean sensor if necessary