beautypg.com

Xylem IQ SensorNet ViSolid 700 IQ (SW) User Manual

Page 34

background image

Measuring

ViSolid

®

700 IQ (SW)

4 - 6

ba76040e02

07/2013

Determining the

Correction factor

Setting the Correction

factor

Note
The Correction factor setting affects each TSS measuring mode and all
calibration data. Also, if there is a change of the measuring mode or
calibration data, the Correction factor is retained.
Therefore, after every change of the settings in the Calibration data
menu, check the Correction factor.

1

Bring the sensor into the measuring position.

2

In the setting table of the TSS sensor, note down the currently
set Correction factor as the value for F

A

.

3

Switch to the measured value display with .

4

When the measured value is stable, read the TSS value,
convert it into the unit (mg/l) if necessary,
and note it down as the value for Sv.

5

Take a sample as close to the same time as the TSS
measurement as possible and as close to the sensor as
possible.

6

Determine the level of total suspended solids of the sample
according to a reference procedure (e. g. gravimetric
according DIN 38414), convert it into the unit (mg/l) if
necessary,
and note it down as the value for S

R

.

7

Calculate the Correction factor.

F

N

= F

A

* S

R

/S

V

8

Set the new Correction factor in the Calibration data menu (see
section 3.5.3).