Xylem IQ SensorNet ViSolid 700 IQ (SW) User Manual

Page 21

background image

ViSolid

®

700 IQ (SW)

Commissioning

3 - 9

ba76040e02

07/2013

Example:

90 ° pipe installation

Fig. 3-6

Total suspended solids sensor in a pipe (90 °)

The following points must be observed for a right-angled installation in
the pipe (Fig. 3-6):

 Rotate the sensor so that the marking on the sensor points in the

direction of the pipe axis

 Select a position where the pipe diameter is as large as possible as

the installation location (see section 3.3.5 D

ISTANCES

FROM

THE

GROUND

AND

WALL

).

Note
In a 90° pipe installation with low levels of total suspended solids
(< 2000 mg/l SiO2 or < 1000 mg/l TSS), the effects of the
measurement environment can have a particularly significant effect on
the measured value. Ensure the optimum conditions of the
measurement environment (see section 3.3.1).

Marking aid

Infrared beam

EBS 700-DU/N

Marking aid
in this direction

ADA-DF 9

1

Connect the SACIQ (SW) sensor connection cable to the plug
head connector of the sensor and screw it tight (see section
3.5.1)
.

2

Attach a marking aid (adhesive strips or similar) in the same
position as the marking on the sensor to the plug head
connector.