Xylem IQ SensorNet ViSolid 700 IQ (SW) User Manual
Page 38
Measuring
ViSolid
®
700 IQ (SW)
4 - 10
ba76040e02
07/2013
Note
To simplify the settings when entering the calibration data,
standardized high resolutions of 1 mg/l are possible. However, it is not
sensible to use these in all cases.
The calibration data entered are evaluated by the system. The entry of
the calibration value pairs can have the following results:
Results after entry of
Calibration data
Note
Information on the contents and structure of the log book and how you
can call it up is given in the L
OG
BOOK
chapter of the IQ S
ENSOR
N
ET
system operating manual.
16
Select the TSS value 1 menu item with <
> and
17
Enter the value for the contents of total suspended solids (TSS
in mg/l) from the reference measurement with <
> and
18
Select the SiO2 value 1 menu item with <
> and
19
Enter the associated SiO2 value (SiO2 in mg/l) measured with
the ViSolid
®
700 IQ (SW) with <
> and
20
Repeat steps 13 - 18 until the required number of value pairs
(between 1 and 8) has been entered.
21
Terminate the entry of the calibration data with Save and quit.
Possible displays
Log book entries
(meaning/actions)
Measured value display
Successful new valid calibration value pairs
were entered for the sensor.
"----"
Calibration value pairs not accepted.
Sensor blocked for measurement.
– Enter the values again making sure that
they are entered in descending order
– Viewing the log book entry.