2 flow direction, 3 sensor angle, Flow direction -3 – Xylem IQ SensorNet ViSolid 700 IQ (SW) User Manual
Page 15: Sensor angle -3
ViSolid
®
700 IQ (SW)
Commissioning
3 - 3
ba76040e02
07/2013
3.3.2
Flow direction
Generally, in flowing media, the measurement window should be
clearly pitched towards the flow (angle of incidence approx. 20 to 45 °).
Exception: If there is a high proportion of foreign bodies with fibrous or
flat profiles such as, e.g. hair, twines or foliage, it can be advantageous
to tilt the sensor in the direction of flow so that the measurement
window is turned away from the flow.
3.3.3
Sensor angle
Fig. 3-1
Effect of the sensor angle on scattering and reflection from the ground and
wall
Note
Scattering and reflection are lowest at a sensor angle of 45° and at a
minimum distance of 10 cm to the ground and walls (see section 3.3.5).
0 °
20 °
45 °
10 cm
Side
view:
Marking
See also other documents in the category Xylem Tools:
- Lab 850 BENCHTOP pH METER (73 pages)
- Lab 850, up to serial number 09470314 (61 pages)
- Lab 860 BENCHTOP pH METER (74 pages)
- Lab 870 BENCHTOP pH METER (71 pages)
- Lab 960 (77 pages)
- Lab 970 (71 pages)
- CG 839 (8 pages)
- CG 842 (44 pages)
- CG 843P (72 pages)
- CG 843 (68 pages)
- CG 853P (76 pages)
- CG 853 (74 pages)
- CG 855 (11 pages)
- HandyLab 100 QuickStart (50 pages)
- HandyLab 100 (66 pages)
- HandyLab 200 QuickStart (50 pages)
- HandyLab 200 (52 pages)
- HandyLab 600 QuickStart (50 pages)
- HandyLab 600 (72 pages)
- HandyLab 680 QuickStart (50 pages)
- HandyLab 680 (120 pages)
- HandyLab 700(EX) (128 pages)
- HandyLab 700 (76 pages)
- HandyLab 780 (164 pages)
- handylab LF 11 (44 pages)
- handylab LF 12 (64 pages)
- handylab OX 12 (60 pages)
- handylab multi 12 (82 pages)
- handylab pH 11 (48 pages)
- handylab pH 12 (68 pages)
- handylab pH LF 12 (80 pages)
- IQ SensorNet FDO 70x IQ (SW) (48 pages)
- MultiLab 4010-2 (112 pages)
- MultiLab 4010-3 (112 pages)
- EBI 40 TK-12 (18 pages)
- TFX 410 (38 pages)
- EBI 20 Serie (20 pages)
- EBI20 TEMPERATURE/HUMIDITY DATALOGGERS (40 pages)
- EBI 20-TF (40 pages)
- TFX 422 (36 pages)
- ST 1000 (42 pages)
- TLC 1598 (2 pages)
- TTX 100/110/120 (28 pages)
- EBI 25 (18 pages)