What to do if ...

ViSolid

®

700 IQ (SW)

6 - 2

ba76040e02

07/2013

Measured value

fluctuating heavily

Measured values too low

Measured values too

high

Cause

Remedy

– There are bubbles of gas in the

medium in front of the
measurement windows

– Check the installation

position of the sensor (see
section 3.3 and section 3.4)

– Signal averaging time too short

for low values of total suspended
solids

– Inhomogeneous measuring

medium

– Increase signal averaging

time

Cause

Remedy

– Measurement window dirty

– Clean the measurement

window (see section 5.2)

Cause

Remedy

– There are bubbles of gas in the

medium in front of the
measurement windows

– Check the installation

position of the sensor (see
section 3.3 and section 3.4)

– Light scattering on the walls

– Check the installation

position of the sensor (see
section 3.3 and section 3.4)

– If necessary, compensate

for any effects that cannot
be cleared by calibration

– Measurement window dirty

– Clean the measurement

window (see section 5.2)