6 what to do if, What to do if -1, 6what to do if – Xylem IQ SensorNet ViSolid 700 IQ (SW) User Manual
Page 43
ViSolid
®
700 IQ (SW)
What to do if ...
6 - 1
ba76040e02
07/2013
6
What to do if ...
Mechanical damage to
the sensor
Display always shows
"0"
TSS display does not
correspond to the TSS
value according to the
laboratory
determination
Display of OFL
Display of "----"
Display of main
parameter
TSS: "----"
Secondary parameter
SiO2: "OFL"
Cause
Remedy
– Return the sensor
Cause
Remedy
– First calibration value pair
incomplete
– Enter the TSS value for the
first calibration value pair
Cause
Remedy
– Correction factor incorrectly set
– Set up the correction factor
again: Correction factor =
TSS value (laboratory) /
TSS value (display)
Cause
Remedy
– Measuring range exceeded
– First calibration value pair
incomplete
– Two identical SiO2 values
entered one after the other
– See log book
– Enter the SiO2 value for the
first calibration value pair
– Enter the value pairs in
descending order
Cause
Remedy
– Measured value invalid
– Incorrect calibration value
entered
– See log book
– Correct the calibration
values and enter them again
Cause
Remedy
– The optical measuring range for
SiO2 is exceeded. The display of
a valid TSS measured value is
not possible.
– See log book
(message code EA6243,
see section 8.1.1)