Commissioning
ViSolid
®
700 IQ (SW)
3 - 14
ba76040e02
07/2013
TSS measuring
mode:
Menu selection:
User calibration
Basic settings
Value pairs 1..3
Value pairs 4..6
Value pairs 7..8
Selection between the use of the basic
settings and the entry of calibration value
pairs.
For the selection of the value pairs, fields
open for the entry of the values
TSS value 1 to TSS value 8.
and the individual values
SiO2 value 1 to SiO2 value 8.
Starting with TSS value 1, determine and
enter the content of total suspended solids in
mg/l TSS, and, in the case of SiO2 value 1,
enter the related measured SiO2 value.
The input precision is 1 mg/l in each case.
Note:
The values must be entered in descending
order.
If this sequence is not kept to, a calibration
error is displayed after leaving the menu. All
the entered values become invalid.
TSS measuring
mode:
Measuring ranges
The setting
AutoRange =
automatic
changeover of the
measuring range
can be selected in all
measuring modes.
0 ... 400.0 mg/l
0 ... 4000 mg/l
0 ... 25000 mg/l
Measuring range for the measuring mode
Mtrx.type1:mg/l TSS
0 ... 400.0 ppm
0 ... 4000 ppm
0 ... 2.500 %
Measuring ranges for the
Matrix type1:% TSS measuring mode
0 ... 4000 mg/l
0 ... 40000 mg/l
Measuring ranges for the
Mtrx.type2:mg/l TSS measuring mode
0 ... 4000 ppm
0 ... 4.000 %
0 ... 40.00 %
0 ... 100.0 %
Measuring ranges for the
Matrix type2:% TSS measuring mode
Setting
Selection/values
Explanation