2 default calibration, Section 4.2.2) – Xylem IQ SensorNet ViSolid 700 IQ (SW) User Manual

Page 31

background image

ViSolid

®

700 IQ (SW)

Measuring

4 - 3

ba76040e02

07/2013

4.2.2

Default calibration

Default calibration for

matrix type 1

The factory calibration curve for matrix type 1 was determined by
measurements of typical activated and return slurries and can be used
for similar applications after adaptation of the Correction factor setting
(see section 4.2.3).

Note
Below the smallest value, the calibration curves are extended to the
zero point and, above the largest value, they are extended to the end
of the measuring range.

Fig. 4-1

Default calibration for matrix type 1

Value pairs

1

2

3

4

5

6

7

8

9

10

Total suspended solids
[1000 mg/l] TSS

17.57 15.55 11.62 8.80 6.21 4.42 3.39 2.40 0.77 0.25

SiO2 value
[1000 mg/l] SiO2

7.16

7.05

6.52 5.85 4.86 3.91 3.22 2.60 1.37 0.61

TS [1000 mg/l]

SiO2 [1000 mg/l ]

25

0,9

20

0,7

15

0,5

10

0,3

5

0,1

0

0

1

2

3

4

5

6

7

8

9

10