beautypg.com

Xylem IQ SensorNet ViSolid 700 IQ (SW) User Manual

Page 32

background image

Measuring

ViSolid

®

700 IQ (SW)

4 - 4

ba76040e02

07/2013

Default calibration for

matrix type 2

The factory calibration curve for matrix type 2 was determined by
measurements of typical decaying slurries and can be used for similar
applications after adaptation of the Correction factor setting (see
section 4.2.3).

Fig. 4-2

Default calibration for matrix type 2

Influences

If there is a low level of total suspended solids (< 2000 mg/l SiO2 or
< 1000 mg/l TSS), the effects of the measurement environment can
simulate a higher content of total suspended solids. The effect of the
measurement environment is minimized by exactly observing the
installation position (see section 3.3.1).

If an optimum installation is not possible due to the structural conditions
at the measuring location (e. g. in narrow pipelines), interference
effects can be compensated by a User calibration (see section 4.2.4).

Value pairs

1

2

3

4

5

6

7

8

9

10

Total suspended solids
[1000 mg/l] TSS

100 59.40 32.00 20.70 14.90 9.97 5.26 2.37 1.48 0.41

SiO2 value
[1000 mg/l] SiO2

7.62

7.16

6.26

5.60

5.00 4.28 3.19 1.73 1.13 0.32

SiO2 [1000 mg/l ]

0

1

2

3

4

5

6

7

8

9

10

TSS [1000 mg/l]

0,9

40

0,7

35

0,5

20

0,3

10

0,1

0

100

0,9

90

0,7

80

0,5

70

0,3

60

0,1

50