YSI DIQ/S 182-MOD System 182 User Manual
Page 98

Relay outputs
System 182
6 - 12
ba76025e01
02/2012
Positive characteristic
curve
The proportional output range begins above the initial value. If the 
proportional range is undercut or exceeded, the selected behavior 
comes into force.
Fig. 6-8
Pulse width output
Fig. 6-9
Frequency output
1
Start value
Measured value with a minimum pulse width or switching 
frequency
2
End value
Measured value with a maximum pulse width or switching 
frequency
0
10
50
90
100
Proportional
band
t
On
t
Of f
50 %
50 %
10 %
10 %
90 %
90 %
Cycle
duration T
Time
Measured value 2
1
Pulse
v [%]
width
0
60
120
Proportional
band
Time
Measured value 2
1
50 % of maximum
switching frequency
Maximum
switching frequency
Switching frequency f [1/min]
Minimum
switching frequency
                    See also other documents in the category YSI Equipment:                                                            
                    - 3100 (50 pages)
 - 3200 (86 pages)
 - 5000 (70 pages)
 - 5200 (148 pages)
 - 5200A (128 pages)
 - 5200A (121 pages)
 - 5400 (127 pages)
 - 5239 Dissolved Oxygen Probe (23 pages)
 - 55 (26 pages)
 - 550A (18 pages)
 - 556 MPS - Quick Start Guide (4 pages)
 - 58 (40 pages)
 - 5905 (22 pages)
 - 6600V2 - User Manual (379 pages)
 - 60 (33 pages)
 - 63 (46 pages)
 - 650 MDS (72 pages)
 - 9300 Photometer (11 pages)
 - 9500 Photometer (143 pages)
 - 9500 Photometer (18 pages)
 - CR 3200 Thermoreactor (42 pages)
 - CR 4200 Thermoreactor (42 pages)
 - CR2200 Thermoreactor (36 pages)
 - DO200 (40 pages)
 - EC300 (40 pages)
 - EC300A (40 pages)
 - EC30A (12 pages)
 - IFL 700 IQ (46 pages)
 - IQ SensorNet 2020 XT Terminal (196 pages)
 - IQ SensorNet FDO Sensor (48 pages)
 - IQ SensorNet MIQ/A-MOD Module (42 pages)
 - IQ SensorNet MIQ/A-PR Module (36 pages)
 - IQ SensorNet MIQ/IC2 Module (30 pages)
 - IQ SensorNet MIQ/IF232 Module (18 pages)
 - IQ SensorNet MIQ/JB Module (18 pages)
 - IQ SensorNet MIQ/JBR Module (20 pages)
 - IQ SensorNet MIQ/PS Module (22 pages)
 - IQ SensorNet MIQ/R6 Module (56 pages)
 - IQ SensorNet NitraLyt Sensor (64 pages)
 - IQ SensorNet SensoLyt Sensor 700 IQ (SW) (46 pages)
 - IQ SensorNet Sensors Encoded Fieldbus Data (12 pages)
 - IQ SensorNet TetraCon Sensor 700 IQ (SW) (32 pages)
 - IQ SensorNet ViSolid Sensor 700 IQ (60 pages)
 - MultiLab 4010-1 (108 pages)
 - MultiLab 4010-2 (138 pages)
 