PROPERTIES menu
4.4.1
Overview
Pressing
g
switches from the measured value display to the
PROPERTIES menu.
In the PROPERTIES menu, you can prompt calibration data and
various information on system components. Moreover, you can switch
on and off the maintenance condition of a sensor.
Fig. 4-7
PROPERTIES menu
Submenus/functions
Maintenance Sensor S
XX
With this function, you switch on and off
the maintenance condition of a sensor (see section 4.4.2)
Sensor status Sxx
Here you can prompt the following data of a sensor (see
section 4.4.3):
– Type, series number and software version
– Data of the last calibration
– Error and info messages
System status
Here you can prompt the following data of the system:
– Series number and software version of the Universal Transmitter
– Status of the operational voltage (
= OK /
= faulty)
– Status of communication (
= OK /
= faulty)
– Current states of interfaces
If the operational voltage or communication are faulty, the flashing
error symbol
h
appears in the header of the measured value
display. The error symbol
h
automatically disappears when the
malfunction has been eliminated.
List of components
Here is a list of all connected sensors.