5 frequency output, Freque, Section 6.4.5 – YSI DIQ/S 182-MOD System 182 User Manual
Page 104

Relay outputs
System 182
6 - 18
ba76025e01
02/2012
6.4.5
Frequency output
Function
The characteristic of the frequency output is laid down in the Start 
value, End value, Frequency (f) min. and Frequency (f) max. settings. 
The fundamentals of the function are described in the introductory 
chapter (see section 6.3.3).
In order to set up the Frequency output function for a relay output, the 
relay output must be linked with a sensor (see section 6.2.1).
Settings
Characteristic curve
If a value is entered for End value that is greater than the Start value, 
the output has a positive characteristic curve.
In order to obtain a negative characteristic curve, a value must be 
entered for End value that is smaller than the value for Start value.
Setting
Selection/Values
Explanation
Measured variable
Main variable
Adjoining variable
Main variable 
designates the 
actual measured 
parameter of the 
sensor (e.g. pH, 
oxygen, etc.).
Adjoining variable 
designates an 
additional 
measured 
parameter (e.g. 
temperature).
Start value
within the 
measuring range 
(sensor-
dependent)
Minimum spacing: 
5 % of the 
measuring range
End value
Frequency (f) min.
0 to 120 1/min
Minimum spacing: 
10 1/min
Frequency (f) max.
Frequency with error
0 to 120 1/min
In case of system 
errors or sensor 
errors (see page 6-
25), the relay 
switches with the 
frequency 
specified.
Action
Open
Close
Relay action