Modbus connection

System 182

7 - 2

ba76025e01

02/2012

7.2

Connecting the Modbus cable

The Modbus cable is connected to the DIQ/S 182-MOD using a 9-pin
D-SUB connector on the top of the housing. As connector on the cable
side, we exclusively recommend one of the following two connectors.

Manufacturer:

Phoenix Contact GmbH & Co. KG
Flachsmarkt 8
32825 Blomberg
Germany
http://www.phoenixcontact.com

Types:

VS-09-PROFB-SC
(Phoenix article no. 1654549, with screwed contacts,
available from YSI under order number 902 888Y)

or:

VS-09-PROFB-SP
(Phoenix article no. 1654345, with spring contacts)

To remove the blind plug, lift the safety bracket with a suitable screw
driver at both clamps on the left and right side (see Fig. 7-1).

Fig. 7-1

Removing the safety bracket

Caution

To prevent the enclosure of the DIQ/S 182-MOD from being
penetrated by water the following must be observed:
Use connectors recommended by YSI only.
When the connector is removed, the connection socket of the

DIQ/S 182-MOD must be closed with the blind plug and secured
with the safety bracket.

Safety bracket