5 behavior of linked relay outputs, 1 behavior in case of error, 2 behavior in non-operative condition – YSI DIQ/S 182-MOD System 182 User Manual
Page 111: Behavior of linked relay outputs -25 6.5.1, Behavior in case of error -25, Behavior in non-operative condition -25
System 182
Relay outputs
6 - 25
ba76025e01
02/2012
6.5
Behavior of linked relay outputs
6.5.1
Behavior in case of error
For linked relay outputs, you can specify the behavior in case of errors.
Depending on the use of the output, the behavior in case of errors is set
in the following menus:
The specified behavior occurs with the following errors.
The system is in the maintenance condition or displays Init, ----, Cal,
Clean, Error or OFL. The meaning of these displays is given in
section 4.2 of this operating manual.
The supply voltage is too low.
6.5.2
Behavior in non-operative condition
An output is non-operative when no function is activated for the output.
The relay contact is open in the non-operative condition.
An output becomes non-operative in case of
Power failure
(As soon as the supply voltage is sufficient again, the non-operative
condition of the outputs ends. The outputs function as specified by
the user again.)
Erasing a link to a sensor
Changing the Measuring mode sensor setting for a linked sensor
Changing the Measuring range sensor setting for a linked sensor
Note
Before editing sensor settings a note appears on the display to inform
you that links will be erased when you change the Measuring mode or
Measuring range sensor setting.
Output
menu
Limit indicator
Behavior at error(see section 6.4.4)
Frequency output
Frequency with error(see section 6.4.5)
Pulse-width output
Error pulse width(see section 6.4.6)