Modbus connection

System 182

7 - 12

ba76025e01

02/2012

Response with all sensor information of the sensor S02:

Modbus response

Byte

Value

Information

Meaning of the value

1

01h

Modbus address of the
DIQ/S 182-MOD

01h --> 1

2

04h

Function

04h --> 4
Read Input Register

3

10h

Number of bytes

10h --> 16
16 Byte (8 registers)

4

02h

Contents of register 9
(HI)
= sensor number

02h (Int 8) --> 2
Sensor number S02

5

02h

Contents of register 9
(LO)
= sensor status

02h (Int 8) --> MEASURE

6

04h

Contents of register 10
(HI)
= sensor model

0401h (Int 16)
--> VisoTurb 700 IQ

7

01h

Contents of register 10
(LO)
= sensor model

8

00h

Contents of register 11
(HI)
= status info

0000h (Int 16)
--> no errors

9

00h

Contents of register 11
(LO)
= status info

10

00h

Contents of register 12
(HI)
= measuring mode

00h (Int 8) --> FNU Turb

11

14h

Contents of register 12
(LO)
= measured value
status

14h (Int 8)
Main measured value (bits
7-4):
1h --> VALID
Secondary measured
value (bits 3-0):
4h --> MISSING