HandyLab 680

Operation

85

ba77041e01

03/2014

Determining the cell

constant

For this calibration procedure, the Type setting must be set to cal . Pro-
ceed as follows to determine the cell constant:

Error

Error

Outside the range
0.450 ... 0.500 cm

-1

Eliminate the error according to
chapter 6 W

HAT

TO

DO

IF

...

Display

Calibration
record

Cell constant [cm

-1

]

1

Connect an IDS conductivity sensor to the meter.

2

In the measured value display, select the conductivity parame-
ter with .

3

Start the calibration with .
The cell constant that was calibrated last is displayed.

4

Immerse the IDS conductivity sensor in the control standard
solution, 0.01 mol/l KCI.

5

Start the measurement with

.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.

6

Wait for the measurement with stability control to be completed
([HOLD][AR]) display or
accept the calibration value with

.
The calibration record is displayed and output to the interface.

7

Switch to the measured value display with

.

15.03.2014 08:00