7 ph value, 1 general information, Ph value 4.7.1 – Xylem HandyLab 680 User Manual
Page 39: General information

HandyLab 680
Operation
39
ba77041e01
03/2014
4.7
pH value
4.7.1
General information
You can measure the following parameters:
pH value [ ]
Sensor voltage [mV]
Note
The sensor connection and the USB-B (device) interface are galvani-
cally isolated. This facilitates interference-free measurements also in
the following cases:
Measurement in grounded test samples
Measurement in one test sample with several sensors connected to
one HandyLab 680
Temperature
measurement
For reproducible pH measurements, it is essential to measure the tem-
perature of the test sample. IDS sensors measure the temperature with
a temperature sensor integrated in the IDS sensor.
Note
The settings for the temperature can be adjusted in the menu for calib-
ration and measurement settings (siehe section 4.7.3).
Preparatory activities
Perform the following preparatory activities when you want to measure:
When operating a sensor without integrated temperature sensor you
have the following options to determine the temperature of the measu-
ring solution:
Temperature measurement via the integrated temperature sensor of
an IDS sensor.
When the measured temperature value is taken from another IDS
sensor the [TP
↑
] status indicator is displayed in the measuring win-
dow. In the measuring window of the IDS sensor providing the mea-
sured temperature value the [TP
↓
] status indicator is displayed.
Manual determination and input of the measured temperature value.
1
Connect the IDS pH sensor to the meter.
The pH measuring window is displayed.
2
If necessary, select the measured parameter with
3
If necessary, calibrate or check the IDS pH sensor.