Xylem HandyLab 680 User Manual
Page 56
Operation
HandyLab 680
56
ba77041e01
03/2014
Continuing with three-
to five-point calibration
Note
After the fifth buffer has been measured the calibration is automatically
finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.
19
Thoroughly rinse the IDS pH sensor with deionized water.
20
Immerse the IDS pH sensor in the next buffer solution.
21
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <
><>.
22
Start the measurement with .
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
23
Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with .
The pH value of the buffer solution is displayed.
24
Set the nominal buffer value for the measured temperature with
<
><>.
25
Accept the calibration value with .
The calibration display for the next buffer appears (voltage dis-
play).
26
If necessary, use
The calibration record is displayed.
or
Continue calibrating using the next buffer with .
15.03.2014 08:00
VariCal