Xylem HandyLab 680 User Manual
Page 42
Operation
HandyLab 680
42
ba77041e01
03/2014
Calibration / Calibration data
storage / Output to USB flash
drive or printer
-
Outputs the calibration data storage to the
interface USB-B (USB flash drive/USB
printer)
Calibration / Calibration data
storage / Output to RS232/USB
-
Outputs the calibration data storage to the
interface USB-B (PC)
Calibration /Buffer
TEC
NIST/DIN
ConCal
...
Buffer sets to be used for pH calibration.
For further buffers and details, see section
4.7.4.
Calibration /Single-point cali-
bration
yes
no
Quick calibration with 1 buffer
Calibration /Calibration interval
1 ... 7 ... 999 d
Calibration interval for the IDS pH sensor (in
days).
The meter reminds you to calibrate regularly
by the flashing sensor symbol in the measur-
ing window.
Calibration /Unit for slope
mV/pH
%
Unit of the slope.
The % display refers to the Nernst slope of
-59.2 mV/pH (100 x determined slope/
Nernst slope).
QSC /First calibration
-
Starts the initial calibration with QSC buffers.
This menu item is only available as long as
no initial calibration was carried out with the
connected IDS sensor.
QSC /Record of first calibration
-
Displays the calibration record of the QSC
initial calibration
QSC /Control calibration
-
Starts the control calibration with QSC buf-
fers.
This menu item is only available if an initial
calibration was carried out with the con-
nected IDS sensor.
Man. temperature
-25 ... +25 ...
+130 °C
Entry of the manually determined tempera-
ture
This menu item is only available if an IDS
sensor without temperature sensor and is
connected.
Menu item
Possible setting
Description