Xylem HandyLab 680 User Manual

Page 42

background image

Operation

HandyLab 680

42

ba77041e01

03/2014

Calibration / Calibration data
storage
/ Output to USB flash
drive or printer

-

Outputs the calibration data storage to the
interface USB-B (USB flash drive/USB
printer)

Calibration / Calibration data
storage
/ Output to RS232/USB

-

Outputs the calibration data storage to the
interface USB-B (PC)

Calibration /Buffer

TEC
NIST/DIN
ConCal
...

Buffer sets to be used for pH calibration.
For further buffers and details, see section
4.7.4.

Calibration /Single-point cali-
bration

yes
no

Quick calibration with 1 buffer

Calibration /Calibration interval

1 ... 7 ... 999 d

Calibration interval for the IDS pH sensor (in
days).

The meter reminds you to calibrate regularly
by the flashing sensor symbol in the measur-
ing window.

Calibration /Unit for slope

mV/pH
%

Unit of the slope.

The % display refers to the Nernst slope of
-59.2 mV/pH (100 x determined slope/
Nernst slope).

QSC /First calibration

-

Starts the initial calibration with QSC buffers.

This menu item is only available as long as
no initial calibration was carried out with the
connected IDS sensor.

QSC /Record of first calibration

-

Displays the calibration record of the QSC
initial calibration

QSC /Control calibration

-

Starts the control calibration with QSC buf-
fers.

This menu item is only available if an initial
calibration was carried out with the con-
nected IDS sensor.

Man. temperature

-25 ... +25 ...
+130 °C

Entry of the manually determined tempera-
ture

This menu item is only available if an IDS
sensor without temperature sensor and is
connected.

Menu item

Possible setting

Description