Xylem HandyLab 680 User Manual
Page 68
![background image](/manuals/310610/68/background.png)
Operation
HandyLab 680
68
ba77041e01
03/2014
Criteria for a stable
measured value
The Stability control function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
4.9.3
Settings for D.O. sensors
(menu or measurement and calibration settings)
Overview
The following settings are possible for D.O. sensors:
Calibration record (display)
Calibration data storage (display)
Calibration interval
FDO Check
Check interval
Salinity correction
Salinity (salinity equivalent)
Settings
The settings are available in the menu for measurement and calibration
settings. To open the settings, display the required parameter in the
measured value display and press the key. After
completing the settings, switch to the measured value display with
Measured
parameter
Time interval
Stability in the time interval
D.O. concen-
tration
20 seconds
Δ
: Better than 0.03 mg/l
D.O. saturation
20 seconds
Δ
: Better than 0.4 %
D.O. partial
pressure
20 seconds
Δ
: Better than 0.8 mbar
Temperature
15 seconds
Δ
: Better than 0.5 °C
Menu item
Possible
setting
Description
Calibration /Cali-
bration record
-
Displays the calibration record of the
last calibration