3 settings for orp measurements, Settings for orp measurements – Xylem HandyLab 680 User Manual
Page 64

Operation
HandyLab 680
64
ba77041e01
03/2014
Criteria for a stable
measured value
The Stability control function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
4.8.3
Settings for ORP measurements
Overview
The following settings are possible for ORP measurements:
Resolution
Automatic stability control
Settings
The settings are made in the menu for measuring settings of the ORP
measurement. To open the settings, display the required parameter in
the measured value display and press the key. After
completing the settings, switch to the measured value display with
Default settings are printed in bold.
Measured param-
eter
Time interval
Stability in the time inter-
val
ORP voltage
15 seconds
Δ
: Better than 0.3 mV
Temperature
15 seconds
Δ
: Better than 0.5 °C
Menu item
Possible set-
ting
Description
mV resolution
0.1
1
Resolution of the mV dis-
play
Reset
-
Resets all sensor settings to
the delivery condition (see
section 4.14.1)