3 settings for ph measurements, Settings for ph measurements – Xylem HandyLab 680 User Manual

Page 41

background image

HandyLab 680

Operation

41

ba77041e01

03/2014

interface without the AutoRead info.

Criteria for a stable

measured value

The Stability control function checks whether the measured values are
stable within the monitored time interval.

The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.

4.7.3

Settings for pH measurements

Overview

The following settings are possible for pH measurements:

 Calibration record (display)

 Calibration data storage (display)

 Calibration interval

 Buffers for calibration

 Single-point calibration

 Unit for slope

 Resolution

Settings

The settings are made in the menu for calibration and measurement
settings of the pH/ORP measurement. To open the settings, display the
required parameter in the measured value display and press the

key. After completing the settings, switch to the mea-
sured value display with .
Default settings are printed in bold.

3

Release the frozen measured value again with or .
The [AR] status display disappears. The display switches back
to the previous indication.

Measured
parameter

Time interval

Stability in the time interval

pH value

15 seconds

Δ

: Better than 0.01 pH

Temperature

15 seconds

Δ

: Better than 0.5 °C

Menu item

Possible setting

Description

Calibration /Calibration record

-

Displays the calibration record of the last cal-
ibration

Calibration /Calibration data
storage

-

Displays the last calibration records (max.
10)