Operation

HandyLab 680

40

ba77041e01

03/2014

4.7.2

Measuring the pH value

Stability control

(AutoRead )

The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values. The display of the measured
parameter flashes until a stable measured value is available.

You can start the Stability control manually at any time, irrespective of
the setting for automatic Stability control (see section 4.4.3) in the Sys-
tem
menu.

Note
You can prematurely terminate the Stability control function manually
with

at any time. If the Stability control function is pre-
maturely terminated, the current measurement data are output to the

1

Perform the preparatory activities according to section 4.7.1.

2

Immerse the IDS pH sensor in the test sample.

3

Select the pH or mV display with .

15.03.2014 08:00

Info

1

Freeze the measured value with .
The [HOLD] status indicator is displayed.

2

Using

, activate the Stability control function
manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized. The progress bar disappears
and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Mea-
surement data meeting the stability control criterion is marked
by AR.