9 dissolved oxygen, 1 general information, Dissolved oxygen 4.9.1 – Xylem HandyLab 680 User Manual
Page 65: General information
![background image](/manuals/310610/65/background.png)
HandyLab 680
Operation
65
ba77041e01
03/2014
4.9
Dissolved oxygen
4.9.1
General information
You can measure the following parameters:
D.O. concentration
D.O. saturation index ("D.O. saturation")
D.O. partial pressure
Note
The sensor connection and the USB-B (device) interface are galvani-
cally isolated. This facilitates interference-free measurements also in
the following cases:
Measurement in grounded test samples
Measurement in one test sample with several sensors connected to
one HandyLab 680
Temperature
measurement
IDS D.O. sensors measure the temperature with a temperature sensor
integrated in the IDS sensor.
Preparatory activities
Perform the following preparatory activities when you want to measure:
Note
The
FDO
®
1100 IDS needs to be calibrated in special cases only. Reg-
ular FDO
®
Checks are sufficient.
1
Connect the D.O. sensor to the meter.
The D.O. measuring window is displayed.
2
Check or calibrate the meter with the sensor.