5 d.o. calibration, D.o. calibration – Xylem HandyLab 680 User Manual
Page 73
![background image](/manuals/310610/73/background.png)
HandyLab 680
Operation
73
ba77041e01
03/2014
4.9.5
D.O. calibration
Why calibrate?
D.O. sensors age. This changes the slope of the D.O. sensor. Calibra-
tion determines the current slope of the sensor and stores this value in
the instrument.
Note
The
FDO
®
1100 IDS D.O. sensor ages so little it does not have to be
regularly calibrated.
To detect changes of the sensor as early as possible, the FDO
®
Check
procedure can be useful (see section 4.9.4).
When to calibrate?
If your evaluation of the FDO
®
Check procedure suggests a calibra-
tion
When the calibration interval has expired
If your accuracy requirements are especially high
Routinely within the scope of company quality assurance measures.
Calibration procedures
Calibration in water vapor-saturated air.
To calibrate the
FDO
®
1100 IDS sensor, use the check and storage
vessel.
Stability control
(AutoRead)
In calibration, the Stability control function (AutoRead) is automatically
activated.
Display calibration data
and output to interface
You can have the data of the last calibration displayed (see section
4.10.6). Subsequently, you can transmit the displayed calibration data
to the interface, e. g. to a PC, with the
Note
The calibration record is automatically transmitted to the interface after
calibrating.
Sample record
Calibration evaluation
After the calibration, the meter automatically evaluates the current sta-
tus of the calibration. The evaluation appears on the display and in the
calibration record.
HandyLab 680
Ser. no. 10139695
CALIBRATION Ox
Calibration date
15.03.2014 16:13:33
FDO 1100 IDS
Ser. no. 10146858
OX 930
10158765
Relative slope
0.98
Sensor
+++