VEGA VEGAPULS 62 (≥ 2.0.0 - ≥ 4.0.0) Modbus and Levelmaster protocol Approval according to LPR radio standard User Manual
Page 66
100 g, 6 ms according to EN60068-2-27 (mechan-
ical shock)
-
Parabolic antenna
25 g, 6 ms according to EN60068-2-27 (mechanical
shock)
Data on rinsing air connection
Pressure max.
6 bar (87.02 psig)
Air quantity without reflux valve, with
-
0.2 bar (2.9 psig)
2.0 m
3
/h
-
0.5 bar (7.25 psig)
3.5 m
3
/h
-
1 bar (14.5 psig)
4.5 m
3
/h
-
1.5 bar (21.76 psig)
5 m
3
/h
-
2 bar (29.0 psig)
5.5 m
3
/h
Thread
G⅛ A
Closure, with
-
Non-Ex
Dust protection cover of PE
-
Ex
Threaded plug of 316Ti
Reflux valve - unmounted (as option with non-Ex version, included in the scope of delivery with Ex
version)
-
Material
316Ti
-
Seal
FKM (Viton), FFKM (Kalrez 6375)
-
for tube diameter
6 mm
-
opening pressure
0.5 bar (7.25 psig)
-
Nominal pressure stage
PN
250
Electromechanical data - version IP 66/IP 67
Cable entry
-
Double chamber housing
l
1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5; 1 x blind stopper
M
16 x 1.5 or optional
6)
1 x plug M12 x 1 for
external indicating and adjustment unit
or:
6)
Only with Ex d version
66
VEGAPULS
62 • Approval according to LPR radio standard
10 Supplement
41722
-EN
-120706