VEGA VEGAPULS 62 (≥ 2.0.0 - ≥ 4.0.0) Modbus and Levelmaster protocol Approval according to LPR radio standard User Manual
Page 53

Code
Text mes-
sage
Cause
Rectification
C
700
Simulation
l Simulation active
l Finish simulation
l Wait for the automatic end
after 60 mins.
The following table shows the error codes and text messages in the
status message "Out of specification" and provides information on
causes as well as corrective measures.
Code
Text mes-
sage
Cause
Rectification
S
600
Unpermissible
electronics
temperature
l Temperature of the elect-
ronics in the non-specified
section
l Check ambient temperatu-
re
l Isolate electronics
l Use instrument with higher
temperature range
S
601
Overfilling
l Danger of vessel overfilling l Make sure that there is no
further filling
l Check level in the vessel
The following table shows the error codes and text messages in the
status message "Maintenance" and provides information on causes as
well as corrective measures.
Code
Text mes-
sage
Cause
Rectification
M
500
Error with the
reset delivery
status
l With the reset to delivery
status, the data could not
be restored
l Repeat reset
l Load XML file with sensor
data into the sensor
M
501
Error in the
non-active li-
nearization
table
l Hardware error EEPROM
l Exchanging the electronics
l Send instrument for repair
M
502
Error in the
diagnosis me-
mory
l Hardware error EEPROM
l Exchanging the electronics
l Send instrument for repair
Out of specification
Maintenance
VEGAPULS
62 • Approval according to LPR radio standard
53
8 Diagnosis, Asset Management and service
41722
-EN
-120706