Fig. 9: Deviating socket dimensions
Tip:
The instrument is optionally available with antenna extension. The
antenna length can thus be selected so that the antenna end protrudes
at least 10 mm (0.4 in) out of the socket.
Align the sensor in liquids as vertical as possible to the product surface
to achieve optimum measurement.
Fig. 10: Alignment in liquids
The mounting location of the radar sensor should be a place where no
other equipment or fixtures cross the path of the microwave signals.
Vessel installations, such as e.g. ladders, limit switches, heating
spirals, struts, etc., can cause false echoes and impair the useful echo.
Make sure when planning your measuring site that the radar sensor
has a "clear view" to the measured product.
In case of existing vessel installations, a false echo storage should be
carried out during setup.
If large vessel installations such as struts or supports cause false
echoes, these can be attenuated through supplementary measures.
Small, inclined sheet metal baffles above the installations scatter the
radar signals and prevent direct interfering reflections.
Sensor orientation
Vessel installations