0 %
100 %
1
2
5
4
3
6
Fig. 14: Configuration bypass
1
Radar sensor
2
Marking of the polarisation plane
3
Instrument flange
4
Distance sensor reference plane to upper tube connection
5
Distance of the tube connections
6
Ball valve with complete opening
Instructions for orientation:
l
Note marking of the polarisation plane on the sensor
l
With threaded fitting, the marking is on the hexagon, with flange
connection between the two flange holes
l
The marking must show to the direction of the tube connections to
the vessel
Instructions for the measurement:
l
The 100 % point must not be above the upper tube connection to
the vessel
l
The 0 % point must not be below the lower tube connection to the
vessel
l
Min. distance sensor reference plane to the upper edge upper
tube connection > 300 mm
l
For the parameter adjustment, select "Application standpipe" and
enter the tube diameter to compensate errors due to running time
shift
l
A false signal suppression with integrated sensor is recommended
but not mandatory
l
The measurement through a ball valve with complete run is
possible
VEGAPULS
62 • Approval according to LPR radio standard
21