Measurement error during filling – VEGA VEGAPULS 62 (≥ 2.0.0 - ≥ 4.0.0) Modbus and Levelmaster protocol Approval according to LPR radio standard User Manual
Page 56
Measurement error during filling
Fault description
Error pattern
Cause
Rectification
4. Measured value re-
mains unchanged du-
ring filling
Level
time
0
l False echoes in the close
range too big or product echo
too small
l Strong foam or spout genera-
tion
l Max. adjustment not correct
l Remove false echoes in the
close range
l Check measurement situation:
Antenna must protrude out of
the socket, installations
l Remove contamination on the
antenna
l Minimize interfering installa-
tions in the close range by
changing the polarization di-
rection
l Create a new false signal sup-
pression
l Adapt max. adjustment
5. Measured value re-
mains in the bottom
section during filling
Level
time
0
l Echo from the tank bottom
larger than the product echo,
for example, with products with
ε
r
<
2.5 oil-based, solvents
l Application parameter "Me-
dium", check Vessel height and
Floor form, adapt if necessary
6. Measured value re-
mains momentarily un-
changed during filling
and then jumps to the
correct level
Level
time
0
l Turbulence on the product
surface, quick filling
l Check application parameters,
change if necessary, e.g. in
dosing vessel, reactor
7. Measured value
jumps towards 0 % du-
ring filling
Level
time
0
l Amplitude of a multiple echo
(vessel top - product surface)
is larger than the level echo
l Check parameter "Application",
especially vessel top, product
type, dished end, high dielect-
ric figure, adapt if necessary
l The level echo cannot be
distinguished from the false
echo at a false echo position
(jumps to multiple echo)
l Remove/reduce false echo:
minimize interfering installa-
tions by changing the polariza-
tion direction
l Chose a more suitable instal-
lation position
8. Measured value
jumps towards 100 %
during filling
Level
time
0
l Due to strong turbulence and
foam generation during filling,
the amplitude of the product
echo sinks. Measured value
jumps to the false echo
l Carry out false signal suppres-
sion
9. Measured value
jumps sporadically to
100 % during filling
Level
time
0
l Varying condensation or con-
tamination on the antenna
l Carry out a false signal sup-
pression or increase false sig-
nal suppression with
condensation/contamination in
the close range by editing
56
VEGAPULS
62 • Approval according to LPR radio standard
8 Diagnosis, Asset Management and service
41722
-EN
-120706