beautypg.com

Measurement error during emptying – VEGA VEGAPULS 62 (≥ 2.0.0 - ≥ 4.0.0) Modbus and Levelmaster protocol Approval according to LPR radio standard User Manual

Page 57

background image

Fault description

Error pattern

Cause

Rectification

10. Measured value
jumps to ≥ 100 % or 0 m
distance

Level

time

0

l Level echo is no longer de-

tected in the close range due
to foam generation or false
echoes in the close range. The
sensor goes into overfill pro-
tection mode. The max. level
(0 m distance) as well as the
status message "Overfill pro-
tection" are outputted.

l Check measuring site: Antenna

must protrude out of the socket

l Remove contamination on the

antenna

l Use a sensor with a more

suitable antenna

Measurement error during emptying

Fault description

Error pattern

Cause

Rectification

11. Measured value
remains unchanged in
the close range during
emptying

Level

time

0

l False echo larger than the

level echo

l Level echo too small

l Remove false echoes in the

close range. Check: Antenna
must protrude out of the socket

l Remove contamination on the

antenna

l Minimize interfering installa-

tions in the close range by
changing the polarization di-
rection

l After removing the false

echoes, the false signal sup-
pression must be deleted.
Carry out a new false signal
suppression

12. Measured value
jumps towards 0 % du-
ring emptying

Level

time

0

l Echo from the tank bottom

larger than the product echo,
for example, with products with
ε

r

<

2.5 oil-based, solvents

l Application parameter "Me-

dium type", check Vessel
height and Floor form, adapt if
necessary

13. Measured value
jumps sporadically to-
wards 100 % during
emptying

Level

time

0

l Varying condensation or con-

tamination on the antenna

l Carry out false signal suppres-

sion or increase false signal
suppression in the close range
by editing

l With bulk solids use radar

sensor with purging air con-
nection

Depending on the reason for the fault and the measures taken, the
steps described in chapter "Setup" must be carried out again or must
be checked for plausibility and completeness.

Should these measures not be successful, please call in urgent cases
the VEGA service hotline under the phone no. +49 1805 858550.

The hotline is also available outside the normal working hours on
seven days a week around the clock.

Reaction after fault rec-
tification

24

hour service hotline

VEGAPULS

62 • Approval according to LPR radio standard

57

8 Diagnosis, Asset Management and service

41722

-EN

-120706