6 set up measured value transmission – VEGA RS232_Ethernet connection User Manual
Page 43
43
9 Measured value transmission to a WEB-VV server
RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62
30325-EN-130701
The MoRoS contains an internal clock for controlling time-dependent
events and dating system messages. Set the time, select the proper
time zone and activate the time synchronization.
Fig. 33: Time synchronisation
9.6 Set up measured value transmission
Select the menu item "Instrument settings - Event list" and press the
button "New". An assistant will now lead you through the configura-
tion.
1. Select in the first window under "Message type" the list entry
"WEB-VV" or "WEB-VV (encoded)".
2. In the second step enter the name or the IP address of the WEB-
VV server. If you’re using the VEGA servers (VEGA Hosting/
WEB-VV VH), enter the URL "web-vv.vega.com" here. If you have
a server of your own (Local Hosting/WEB-VV LH), you can get
the server name or the IP address from your responsible network
administrator.
3. In the last step, define when or how often the measured values
are transmitted. Please remember that costs arise with every
measured value transmission, depending on the type of connec-
tion.
4. Transfer your settings to the signal conditioning instrument.
After completing the configuration, you should start a test transmis-
sion to the WEB-VV server. Then contact the WEB-VV administrator.
He will allocate the new instrument network accordingly and release
the measured values for your inspection.
Time