Relay outputs, Settings in the "extended menu – VEGA VEGAMET 614 User Manual
Page 33

VEGAMET 614
33
Settings in the "Extended menu“
Relay outputs
In the menu item "
prop. to
“, you choose a
parameter to which the relay reacts. You can
choose between percent and volume ("prop.
to volume“ is useful when under "Conditioning
- Linearisation curves“, for example, a spheri-
cal tank or a user-programmable lin. curve is
selected). In the next menu item, you can
select the suitable
Unit
. In the menu item
"
Mode
“, the mode of the relay can be se-
lected. The individual relay modes are de-
scribed on the following pages.
If mode "Rising tendency“ or "Falling ten-
dency“ is selected, the menu items "Devia-
tion“, "Dev. per time“, "Cycle time“ and
"Number of cycles“ appear.
Param.
TAG-
No. 1
TAG -
No. 1
%
xx,x
Adjust-
ment
Signal
condit-
ioning
Out-
puts
Cur-
rent
out-
puts
Volt
outputs
Relay
outputs
Relay
output
1
Relay
output
2
prop.
to
Per-
cent
Unit
0,0%
Mode
Over-
fill
protec.
Low
%
0,0
Failure
mode
off
Add’l
func-
tions
Cycle
time
s
1
Number
of
cycles
10
like relay output 1
High
%
100,0
Devia-
tion
s
1,0
Devia-
tion
per
time
Switch-
ing
delay
t off
s
0
Volume
Height
Pressure
Dry run protection
Switching window on
Switching window off
Rising tendency
Falling tendency
0,00 %
hold
t on
s
0
With all other modes, the menu items "Low“
and "High“ appear. There you can determine
the switching points for the relay.
The menu item "
Failure mode
“ always ap-
pears. "Failure mode off“ causes the relay to
deenergise when a failure (e.g. sensor
shortcircuit) occurs. "Failure mode hold“
causes the actual relay status to be main-
tained when a failure occurs.
Under "Add’l functions“ you can adjust a
switch on delay (t on) and a switch off delay
(t off) of the relay of max. 600 s (with the relay
modes "Rising tendency“ or "Falling ten-
dency“ no switching delay can be set).