beautypg.com

4 adjustment – VEGA VEGAMET 614 User Manual

Page 30

background image

30

VEGAMET 614

Settings in the "Extended menu“

7.4 Adjustment

Param.
TAG-
No. 1

Adjust-
ment

with
medium

w.o.
medium

Min-
adjust
at %

0,0

Max-
adjust
at %

100,0

Adjust-
ment

mA

Offset
correc-
tion

0%
at

mA

4,000

100%
at

mA

20,000

Sensor
pres-
sur’d ?
OK ?

Offset
corr.
Now !
OK ?

TAG -
No. 1
%
xx,x

The

adjustment with medium

corresponds

to the procedure in the reduced menu (see
appropriate chapter).

- if the medium is changed later on (medium

with another density value or DK value)
and no new adjustment can be made
or

- if the MET indication and the DIS outputs

relate to height in meters

note the following:
In applications with hydrostatic pressure
transmitters, the adjusted density value (see
menu "Parameter, Signal conditioning“) must
correspond to the value of the medium. In
applications with capacitive electrodes, the
adjusted DK value must correspond to the
value of the medium.

For the

adjustment without medium,

two

anticipated values for level, gauge etc. must
be adjusted corresponding to 0 % and 100
% must be entered.

The menu item "Offset correction“ is activated
when hydrostatic pressure transmitters are
connected. The offset correction should be
carried out after installation of the sensor as
the sensor values can differ slightly due to
varying installation positions (vertical or hori-
zontal installations). With the offset correction,
the meas. data of the unpressurised sensor
(vessel empty) are detected as correction
values and taken into account in subsequent
measurements.