beautypg.com

5 dimensions – VEGA VEGAMET 514V User Manual

Page 8

background image

8

VEGAMET 514V

Product description

Electrical protective measures

Protection
- unassembled

IP 00

- mounted into

carrier BGT 596
- front side completely equipped

IP 40

- upper and lower side

IP 20

- wiring side

IP 00

- mounted into

housing type 505 or type 506
- terminal side

IP 20

- housing general

IP 30

Protection class

II (in housing type 505)

Overvoltage category

II

Electrical separating measures

Reliable separation acc. to
VDE 0106, part 1 between

power supply, fail safe and level relay
and measuring data inputs

- reference voltage

250 V

- isolation resistance

2.3 kV

Galvanic isolation

among the relay outputs

- reference voltage

250 V

- isolation resistance

1.4 kV

Potential separation

among DISBUS and outputs

- reference voltage

50 V

- isolation resistance

0.5 kV

Common reference potential at

voltage and current output,
correction signal output

Mechanical data

Series

module unit for carrier BGT 596
or housing type 505

Dimensions

B = 25.4 mm (5 TE), H = 128.4 mm, T = 162 mm

CE conformity

VEGAMET 514V signal conditioning instrument meets the protective regulations of EMC
(89/336/EWG) and NSR (73/23/EWG). Conformity has been judged acc. to the following
standards:
EMC

Emission

EN 50 081 - 2: 1992

Susceptibility

EN 50 082 - 2: 1995

NSR

EN 61 010 - 1: 1993

1.5 Dimensions

162

100

5,5

15

128,4

25,4

5 TE

Multiple plug

Circuit board 100 x 160 x 1.5
European size