Settings in the "reduced menu“
Name measurement loop
TAG
name
T
AG -
No. 1
TAG
name
T
A
G -
No. 1
TAG
name
Ves.4
Hall
3
Ves.4
Hall3
%
27,8
6.1 Configuration of measurement
loop
Choose sensor type
By default, VEGAMET 514 was configured for
connection to a hydrostatic sensor. If you are
using another sensor, first of all you have to
carry out a "Reset TAG 1“ (confirm in the
menu item "Reset Now! OK?“ with the [OK]
key). Then move to the menu item "Sensor
type Hydrostatic“, there you choose the
sensor type with the keys [+] or [–] and
confirm with [OK].
TAG-
No. 1
%
xx,x
Param.
TAG-
No. 1
Confi-
gura-
tion
TAG
name
TAG -
No. 1
Sensor
type
Hydro-
static
Sensor
opti-
mize
Meas.
value
indication
TAG-
No. 1
%
xx,x
Param.
TAG-
No. 1
Confi-
gura-
tion
Add’l
func-
tions
Pass-
word
off
Lang-
uage
Eng-
lish
TAG 1
to de
fault
Sensor
type
Hydro-
static
Reset
TAG 1
Delete
TAG
1?
Reset
OK ?
Reset
Now!
OK ?
Pulse-Echo
Pulse-Radar
You can modify the measurement loop name
"TAG-No. 1“, if you want a more useful desig-
nation.
Two lines with six characters each are avail-
able. To do this, move to the menu item "TAG
name …“.
Push [+] or [–] once. The first char-
acter of the measurement loop name,
in this example the letter T, flashes
(editing mode). Continue pushing the
[+] or [–] key, by this you scroll through the
alphabet, through all the numbers and
through a list of characters.
When you have reached the re-
quested character, push [–>]. The
next character will be activated and
flashes. Scroll again with [+] and [–]
to the requested character. With [–>] you
reach again the next character.
When you have finished the last posi-
tion of the name and you have "writ-
ten" the name of the measurement
loop in this way, you have to push
[OK] to save the setting. When you quit the
menu item without saving, the name of the
measurement loop is again "TAG-No. 1“.
In the display of measured value and
in all other menu items in which the
measurement loop name is stated,
you will now find the new designation.
6 Settings in the "reduced menu“
It is only possible to modify
the sensor type, when a
"Reset TAG 1“ was carried
out before.