VEGASON 54V … 56V

Setup - Adjustment with the PC on VEGALOG

Parameter adjustment

In the menu „

Instrument data/Parameter

adjustment“ you carry out all important sen-
sor settings.

Adjustment

• Choose the menu „

Instrument data/Param-

eter adjustment“ and then the sensor for
which you want to carry out the parameter
adjustment.

In the header of the opening menu window,
you see the previously entered measurement
loop name and the measurement loop de-
scription. If you have configured only one
sensor on the signal conditioning instrument,
only a choice of one sensor will be offered.

• First choose „

Adjustment“.

• Click in the menu window „

Adjustment“ to

Min/Max-Adjustment“.

You can carry out the min./max. adjustment
"with medium“ (adjustment by means of the
actual level) or

"without medium“ (without

taking the actual level into consideration, i.e.
with empty vessel).

Generally, you will carry out the adjustment
without medium. Then you are not concerned
about the actual vessel filling during adjust-
ment. When you want to carry out the adjust-
ment with medium, you have to carry out the
min. adjustment with emptied (also partly
emptied) vessel and the max. adjustment
with filled (also partly filled) vessel. It is con-
venient and quick to carry out the adjustment
without medium, as shown in the example.

• Choose "

no (adjustment without medium)“.

• Choose if you want the carry out the ad-

justment in

meters (m) or in feet (ft).

• Enter a distance for the upper and lower

level and the volume in % corresponding to
each distance.

In the example, the 0 % filling is at a product
distance of 3.400 m and the 100 % filling at a
product distance of 0.500 m.

Note:

The sensor can only detect levels within the
defined operating range. For the detection of
levels outside the operating range, e.g. new
adjustment, it is necessary to correct the
operating range in the menu "

Sensor optimi-

sation/Meas. environment“.