beautypg.com

Dished tank ceiling, Open vessels – VEGA VEGASON 54V…56V User Manual

Page 31

background image

VEGASON 54V … 56V

31

Open vessels

Min. meas.
distance

Dished tank ceiling

On dished tank ceilings, please do not mount
the instrument in the centre, but approx.

1

/

2

vessel radius from the centre. Dished tank
ceilings can act as paraboloidal reflectors. If
the transducer is placed at the focal point of
the parabolic ceiling, the transducer receives
amplified false echoes. The transducer
should be mounted outside the focal point.
Amplified echoes caused by parabolic sur-
faces are thereby avoided.

Mounting and installation

Reference plane

< 100 mm

VEGASON 54 on dished vessel ceiling; the state-
ments are also valid for VEGASON 55

Reference plane

1

/

2

vessel radius

< 400 mm

VEGASON 56 on dished tank ceiling

Reference plane

1

/

2

vessel radius

Open vessels

On open vessels, use of instruments on an
extended mounting bracket is recom-
mended. Mount the low-weight sensor onto
such a bracket and ensure a sufficient dis-
tance to the vessel wall.

Min. meas.
distance

Reference plane