• Choose one of the two inputs of the VEGA-

MET signal conditioning instrument (VEGA-
MET 514V has only one sensor input) and
click to „

OK“.

After a few seconds, the menu window „

Cre-

ate new measurement loop - Sensor configu-
ration“ opens.

• Click in the menu window „

Create new

measurement loop - Sensor configuration“
to „

Sensor coordination“. The menu win-

dow „

Sensor coordination“ opens.

• Click to „

Sensor search“.

• Then click to „

Input“.

• Choose the serial number of the sensor to

which you want to allocate e.g. input 1.

• Confirm with „

OK“.

• Click in the menu window „

Sensor coordi-

nation“ again to „

OK“.

• Choose „

Standard level measurement“

and „

no options“.

• Click to „

Continue“ and the menu window

New application - Select measurement

loop“ opens.

• Choose the parameter (

„level measure-

ment“, gauge or distance) and the sensor
type (

„Pulse Echo“ for ultrasonic).

• Click to „

Continue“.

Create new measurement loop

• Choose the menu

„Configuration/Measure-

ment loop/New“ and you are in the menu
window „

Create new measurement loop -

Application“.

Setup - Adjustment with the PC on VEGAMET