18

VEGASON 54V … 56V

Technical data

Characteristics

1)

(typical values under reference conditions, all statements relate to the nominal measuring
range)

Min. measuring range

(between empty and full adjustment)

> 20 mm (recommended > 50 mm)

Ultrasonic frequency (at 20°C)
- VEGASON 54

30 kHz

- VEGASON 55

18 kHz

- VEGASON 56

10 kHz

Measuring intervals
- VEGASON 54

1.0 s

- VEGASON 55

1.5 s

- VEGASON 56

2.0 s

Beam angle at -3 dB emitted power
- VEGASON 54

- VEGASON 55

- VEGASON 56

Influence of the process temperature

1.8 %/10 K, however is compensated by an

integral, dynamic temperature detection
system in the transducer

Influence of the process pressure

negligible within the approved sensor pressure
range

Ambient conditions

Ambient temperature (housing)

-20°C … +60°C

Process temperature (transducer)
- VEGASON 54, 55

-40°C … +80°C (StEx: -20°C … +75°C)

- VEGASON 56

-40°C … +150°C

- storage and transport temperature

-40°C … +80°C

Vessel pressure max. (gauge pressure)
- VEGASON 54

version A

1.5 bar (flange version)

version B … C

0.5 bar

version D

3 bar

- VEGASON 55

version A

1.5 bar (flange version)

version B … C

0.5 bar

version D

3 bar

- VEGASON 56

version A

3 bar (flange version)

version B … C

0.5 bar

version D

3 bar

Protection
- sensor

IP 67

- transducer, process

IP 68

Protection class
- two-wire sensor

II

- four-wire sensor

I

Overvoltage category

III

Self-heating
at 40°C ambient temperature
- sensor

45°C

- transducer, process

55°C