beautypg.com

VEGA VEGASON 54V…56V User Manual

Page 62

background image

62

VEGASON 54V … 56V

• Quit the menu with

"OK“.

• Confirm with

"OK“ and your individual lin-

earisation curve is saved in the sensor.

Again in the menu window "

Conditioning“, you

can enter with the menu item "

Integration

time“ a measured value integration. This is
recommended for agitated product surfaces,
to prevent the measured value indication and
output from changing constantly.
As a standard feature, an integration time of
0 seconds is preset.

• Quit the menu with

"OK“., you are again in

the menu window "

Instrument data param-

eter adjustment“.

• Quit the menu window with

"OK“.

Setup - Adjustment with the PC on VEGAMET

The sensor then outputs the actual filling
volume by means of the adjusted vessel
dimensions.

There is a linear interpolation between the
linearisation points.

• Click to

"OK“ and you are again in the

menu window "

Tank calculation“.

• Again click in the menu window "

Tank cal-

culation“ to

"OK“ and you are in the lineari-

sation menu.

Here the calculated linearisation curve is
again outputted. The volume information
under "

Linearised“ now corresponds no

longer to the calculated volume of the tank
calculation program. Why?

In the menu "

Scaling“ (Instrument data/Condi-

tioning/Scaling) you entered that at 0 % filling
there are 45 liters in the tank and at 100 %
filling 1200 liters. The geometry of the calcu-
lated cylindrical tank is reduced to a size that
has a volume of only 1200 liters. The lineari-
sation curve of the calculated vessel is there-
fore modified to fit the volume data you
entered in the menu "

Scaling“.

If the true content of the calculated vessel
should be outputted, the volume that was
determined in the tank calculation program
must be entered in the menu "Scaling“.