beautypg.com

VEGA VEGASON 54V…56V User Manual

Page 19

background image

VEGASON 54V … 56V

19

Technical data

Ex technical data

Classification

ia intrinsically safe (in conjunction with a safety
barrier or separator)

Temperature class (permissible ambient
temperature around the transducer when
used in Ex areas)
- T6

42°C

- T5

58°C

- T4

60°C

- T3

60°C

Ex approved in category and zone
- VEGASON 56

ATEX

Zone 1 (II 2G)

IEC, CENELEC, PTB

Zone 1

- VEGASON

ATEX

Zone 21/22 (II 2D/3D)

IEC, CENELEC, PTB

Zone 10/11

Process fittings

VEGASON 54

G 1 A, 1-11.5 NPT, DN 50, DN 80, DN 200

VEGASON 55

G 1 A, 1-11.5 NPT, DN 50, DN 80, DN 250

VEGASON 56

G 1 A, 1-11.5 NPT, DN 50, DN 80, DN 200

Note:
Generally an adapter flange is necessary for G 1 A, 1-11.5 NPT, DN 50, and DN 80.

Connection cables

Two-wire sensors

supply and signal in two separate cables

Electrical connection

spring-loaded terminals, wire cross-section

generally 2.5 mm

2

- option

screw connection

Ground connection

max. 4 mm

2

Transducer cable
- VEGASON 54, 55

5 … 300 m (cable diameter 7.2 … 7.6 mm)

- VEGASON 56

5 … 300 m (cable diameter 9.5 … 9.9 mm)

Cable entry for signal and supply cable
with spring-loaded terminals
- plastic housing (PBT)

2 x M20 x 1.5 (cable diameter 5 … 9 mm)
or 2 x

1

/

2

“ NPT (cable diameter

3.1 … 8.7 mm or 0.12 … 0.34 inch)

- Aluminium housing

2 x

1

/

2

“ NPT (cable diameter 3.1 … 8.7 mm

or 0.12 … 0.34 inch)

1)

Similar to DIN 16 086, references conditions acc. to IEC 770;
Temperature 15°C … 35°C; moisture 45 % … 75 %; air pressure 860 mbar … 1060 mbar