Setup - Adjustment with the PC on VEGALOG
The adjustment program VEGA Visual Ope-
rating (VVO), called in the following VVO,
gets into contact with the connected sensor
and indicates after a few seconds if and with
which sensor a connection exists.
Note:
If you make no connection with the sensor,
check the following:
- Is the processing system being supply
with supply voltage?
- Do you inadvertently use a wrong RS 232
cable instead of the VEGA RS 232 connec-
tion cable?
If VVO (adjustment software) gets in contact
with the processing system for the first time,
you will be asked if you want to transfer the
data from the processing system to the PC.
• Click to „
Yes“.
If connecting the VVO to a VEGALOG from
which data has already been saved, you are
asked if the saved data should be trans-
ferred from the PC to VEGALOG or the data
from VEGALOG to the PC.
• Click to
OK and you are in the main
menu window.
The preadjusted user identification can be
modified at a later time in the menu „
Configu-
ration/Program/User access“.
Independent whether you set up a measur-
ing system (unit consisting of sensor and
VEGAMET signal conditioning instrument or
VEGALOG processing system) with the
adjustment software VVO or with the signal
conditioning instrument, the adjustment pro-
cedure is always the same:
- first of all crease a measurement loop and
configure a measuring system in the menu
„
Configuration“ and then
- carry out the parameter adjustment of the
sensor in the menu „
Instrument data“.