Adjustment with MINICOM or VEGAMET

Indication and adjustment surfaces

ESC

OK

-

+

1

2

on

100

%

CONNECT

515 V

Acc. to the menu point, change
value or choose out of a list

Choose menu window or shift
flashing cursor

Save the adjusted value or
change to the lower menu

Branch, i.e. jump to the
lower menu with [OK]

Interrupt the
adjustment or jump to
the upper menu

Display,
indication of
- measured value
- menu point
- parameter
- value

Analogue
LED-indication
(0 … 100 %)

6.3 Adjustment with MINICOM or

VEGAMET

Beside the PC, VEGAPULS 51 V … 54 V
radar sensors can be also adjusted

- with the small, detachable

adjustment module MINICOM

- or with the signal conditioning

instrument.

All adjustment options are available for the
adjustment with signal conditioning

signal conditioning

signal conditioning

signal conditioning

signal conditioning

instrument VEGAMET

instrument VEGAMET

instrument VEGAMET

instrument VEGAMET

instrument VEGAMET like with PC. The
adjustment just differs in the demonstration
however not in the function.

With the adjustment module MINICOM

adjustment module MINICOM

adjustment module MINICOM

adjustment module MINICOM

adjustment module MINICOM all

sensor relevant adjustments (adjustment,
meas. range, meas. conditions, sensor-
indication-scaling or false echo memory) are
available.
However not possible are the adjustment
steps relating to the configuration of the
signal conditioning instrument VEGAMET or

the processing system VEGALOG and their
signal processing (configuraton of inputs and
outputs, linearisation curves, simulation…).

The adjustment of VEGAMET and MINICOM
is identical. They are adjusted with 6 keys. A
small display shows you beside the
measured value, in short words a message
on the menu point or on a figure of a menu
adjustment.
The information quantity of the small display
however cannot be compared with the
adjustment program VVO, however in
conjunction with the following menu plan of
MINICOM

MINICOM

MINICOM

MINICOM

MINICOM and VEGAMET 515 V

VEGAMET 515 V

VEGAMET 515 V

VEGAMET 515 V

VEGAMET 515 V you will

have no problems with the adjustment. Some
users however will find the work with the
small adjustment field quicker and more
direct to make than with the PC.
The menu plan to VEGAMET 514 V is
provided in the operating instruction of
VEGAMET 514 V.

Signal conditioning
instrument VEGAMET

Adjustment module
MINICOM

ESC

OK

Tank 1
m (d)
12.345