VEGA VEGAPULS 51V…54V User Manual

Page 62

background image

62

VEGAPULS 51 V … 54 V

Backup

• Click to

"

Services/Backup/Sensors

Services/Backup/Sensors

Services/Backup/Sensors

Services/Backup/Sensors

Services/Backup/Sensors"

In the menu window "

Backup" the sensor with

the serial number is displayed. You can save
the sensor individually or in groups with all
adjustments in a directory of your choice on
your PC. In addition you can add a short note
to each backup.

You also save the adjustments of the signal
conditioning instrument VEGAMET:

Saved sensor data can be transferred later
to other sensors. If you have e.g. a system
with several of the same storage vessels and
identical sensors, it is sufficient to configure
one sensor and then transfer to the other
sensors.

• Choose the menu """""

Services/Restore

Services/Restore

Services/Restore

Services/Restore

Services/Restore

configuration/Sensors

configuration/Sensors

configuration/Sensors

configuration/Sensors

configuration/Sensors"""""

In this menu window you choose in the yellow
window cutout a sensor (or a signal
conditioning instrument) which should be
overwritten by the adjustments of another
sensor.

Choose in the field "

Selection of backups" the

serial number of the sensor of which you
want to transfer the adjustments. With
"

Restore to" you transfer these sensor

adjustments to the sensor you have chosen
in the yellow window cutout.

Set-up

-

Adjustment with the PC on VEGAMET