VEGA VEGAPULS 51V…54V User Manual
Page 20
20
VEGAPULS 51 V … 54 V
Ex-technical data
(note approval documents)
Flame proofing
ia (intrinsically safe in conjunction with
a safety barrier or separator)
Classification
EEx ia IIC T6
Perm. housing ambient temperature
T6: 45°C
T5: 58°C
T4/T3: 60°C
Temperature class (permissible ambient temperature on the antenna system)
- T6
80°C
- T5
95°C; type 51: 80°C
- T4
130°C; type 51: 80°C
- T3
150°C; type 51: 80°C; type 52: 130°C
Ex-approved in category or zone
- EC-type approval
Type 5*V Ex: Zone 1 (II 2 G)
Type 5*V Ex 0: Zone 0 (II 1 G)
- conformity certificate
Type 5*V Ex: Zone 1
Type 5*V Ex 0: Zone 0
Process connections
VEGAPULS 51, 52
G 1
1
/
2
A, 1
1
/
2
" NPT (rod antennas on plastic
or StSt thread)
VEGAPULS 53
DN 50, DN 80, DN 100, DN 150 (rod antennas)
VEGAPULS 54
DN 50, DN 80, DN 100, DN150
ANSI 2", 3", 4" and 6"
(up to DN 100 or 4" mounting on surge pipe)
Connection lines
Two-wire sensors, supply and signal via one two-wire line
Line resistance max. 15
Ω per wire or 1000 m cable length
Cross-section area of conductor
generally 2,5 mm
2
Earth connection
max. 4 mm
2
Cable entry
2 x M20 x 1,5 (cable diameter 5 … 9 mm)
Materials
Housing
PBT (Valox)
Flange / process connection
- VEGAPULS 51
PVDF or StSt
- VEGAPULS 52
PVDF or StSt
- VEGAPULS 53, 54
1.4571
Antenna
- VEGAPULS 51
PP or StSt / PPS
- VEGAPULS 52
PTFE or StSt / PTFE
- VEGAPULS 53
PTFE
- VEGAPULS 54
1.4571, 1.4071
Flange coating (only VEGAPULS 53)
PTFE
Technical data