VEGAPULS 51 V … 54 V

Sensor optimisation

In the menu "

Sensor optimisation" you can

carry out special optimizing adjustments of
the sensors and can optimize e.g. by means
of the echo curve the installation position of
the sensor.

Meas. environment

• Choose in the main menu window the menu

"

Instrument data/Parameter adjustment

Instrument data/Parameter adjustment

Instrument data/Parameter adjustment

Instrument data/Parameter adjustment

Instrument data/Parameter adjustment"

• Choose in the menu window

"Instrument

data parameter adjustment" the menu point
"

Sensor optimisation

Sensor optimisation

Sensor optimisation

Sensor optimisation

Sensor optimisation", and click to

"

Sensor A

Sensor A

Sensor A

Sensor A

Sensor A"

Set-up

-

Adjustment with the PC on VEGAMET

• Choose under

"Sensor No." "A

A

A

A

A" and under

"Parameter" "Scaled

Scaled

Scaled

Scaled

Scaled".

• Choose under "

Scaling for sensor display"

e.g. litres.

• Enter the meas. distance in meter which

you have entered for the min./max.-
adjustment and the appropriate litre values
corresponding to the min.-value and the
max.-value. In the example 45 litres and
1200 litres.

• Click to

"Save

Save

Save

Save

Save"

• Click in the window "

Sensor-Display" to

"Quit

Quit

Quit

Quit

Quit"

• Click in the window "

Display of measured

value" to "Quit

Quit

Quit

Quit

Quit"

You are again in the menu window "

Outputs"

• Click to

"Quit

Quit

Quit

Quit

Quit"

You are again in the menu window "

Instru-

ment data parameter adjustment".