beautypg.com

Veris Industries TURBO SERIES Recordal Install User Manual

Page 6

background image

carefully pressing (with flat surface of blade screwdriver) on bottom of
tube located inside rear nose cone. Press to release detent lock,
approximately 1/4" vertical travel, then lift tube out from top of cage
insert. Lift transmission shaft assembly up and out of bottom bearing,
then slide shaft downward and diagonally out (towards you) to remove
from nose cone/cage assembly. Finally, holding cage firmly with one
hand, rotate nose cone clockwise with other hand, then carefully lift off
cage. Reinstallation is reverse of removal.

Refer to Section 4-8 of this manual to correctly reassemble compo-
nents.

4-6. CALIBRATION MECHANISM

To disassemble the calibration ring assembly, the following proce-
dures are required (see Figures 3a and 3b). Refer to Sections 4-2
through 4-5 above. Remove the calibration shaft seal plug and
unscrew the calibration shaft lock screw from the cover plate. Remove
the calibration shaft thrust washer and calibration shaft. The calibra-
tion shaft can be pressed out of the cover plate from below. The
calibration shaft O-ring may be removed if a leak exists at this point in
the disassembly of the meter. If replaced, the O-ring should be coated
slightly with a silicon grease before insertion. Clean all parts as
required.

To gain access to the calibration ring, the straightening vane/nose
cone must be removed. See Section 4-5 to remove. To remove
calibration ring from nose cone assembly, align the ring slots with
bayonettes of nose cone assembly located on inner ring of calibration
ring. To remove, carefully lift the calibration ring from the nose cone
assembly. Reinstallation is the reverse of this procedure.

Page 6

Figure 3b. Register and Measuring

Element Assembly, 6"

MAGNET

CARRIER

CAGE

INSERT

HEAD

O-RING

COVER

PLATE

NOSE CONE

SETSCREW

TRANSMISSION

SHAFT GEAR

PROTECTION

TUBE

EYEBOLTS

HEAD BOLT

SEAL PLUG

LOCK SCREW

THRUST WASHER

O-RING

CALIBRATION GEAR

CALIBRATION

SHAFT

CALIBRATION

DRIVE PLATE

RETAINING

RING

Figure 4a. Calibration Ring/Linkage Assembly, 1 1/2" - 4"

CALIBRATION

LINKAGE

ARROW

TIP

CAGE

INSERT

NOSE
CONE

TOP SIDE

SMALL HOLE

WITH

CHAMFER UP

LARGE HOLE

WITH

CHAMFER

DOWN

INTO CAGE

THIS END

CALIBRATION

RING

PIN

4-4. MEASURING ELEMENT ASSEMBLY

Remove the O-ring and cage seal. Check for damage and clean or
replace prior to reassembly.

NOTE

A cage seal is installed in the 1 1/2" - 6" Recordall

®

Turbo

Series so that the entire water volume flows through the
measuring element.

To remove measuring element from cover, remove calibration shaft
seal plug and lock screw (see Figures 3a and 3b). Place cover register
side down on table or flat surface. Lightly tap or press calibration shaft
from wet side out of bore in cover (1 1/2" - 4" only). Care should be
taken not to strike or bend the calibration linkage.

The measuring element insert can be removed from the cover plate for
service or replacement by removing the cage set screw (see Figures
3a and 3b, quantity 4 for 6" size. Holding the cover plate in one hand,
rotate the thermoplastic element counter clockwise until the bayonette-
like tabs are aligned with open areas of the cover plate (1 1/2" - 4" only).
Remove element from cover plate.

4-5. STRAIGHTENING VANES/NOSE CONE

The straightening vanes are an integral part of the upstream and
downstream nose cone assemblies (see Figures 3a and 3b). To gain
access, the measuring element must be removed from cover as
outlined in Section 4-4 above.

Removal of upstream nose cone provides access to rotor, calibration
ring and transmission assembly. To remove nose cone, unscrew the
upstream nose cone setscrew. For 1 1/2" - 4", slide calibration linkage
into cage insert stop. Take hold of the nose cone/straightening vanes
and turn clockwise, thereby disengaging the calibration linkage from
the tab on the calibration ring and unlocking bayonettes of nose cones
from cage (see Figure 4a). If tab of calibration ring has not disengaged
linkage, rotate ring carefully by pushing struts at interior of element
(between straightening vanes and rotor). Pull the nose cone assembly
out from the measuring element insert. For 6", remove transmission
gear by lifting off shaft, noting alignment of "D" shaped bore with flat

that holds down calibration drive plate, located under magnet carrier
magnet (see Figures 3b and 4b). Remove drive plate. Rotate
calibration ring until pin clears slot of cage. Take hold of the nose cone/
straightening vanes and turn clockwise. Pull the nosecone assembly
out from the measuring element insert. Remove rotor assembly.

Rear nose cone removal requires transmission shaft removal (see
Figures 3a, 3b, 4a and 4b). First, remove strap over gear (1 1/2" - 4"
only). Gear is removed from shaft by simply lifting off. Note shaft has
a flat that matches with "D" shaped bore in gear (required to align for
assembly). Remove nose cone set screw. Remove protection tube by

on shaft (required to align for assembly). Next remove retaining ring

NOSE CONE/

STRAIGHTENING

VANE ASSEMBLY (2)

TRANSMISSION

SHAFT ASSEMBLY

BEARING

CALIBRATION

RING (REAR)

CAGE SEAL

CALIBRATION

RING (FRONT)

CAGE

SETSCREW (4)

WASHER (4)

ROTOR