TROUBLESHOOTING

24

Scale counts,
but not
correctly

Check that the scale cable

is securely connected to

the readout.

Loose cable

Check that the scale cable

is connected to the correct

input on the back of the

readout.

Wrong input

Check the scale and read-

ing head for proper instal-

lation (refer to scale instal-

lation instructions).

Scale not installed
properly

Scale won’t
count

Check the scale and read-
ing head for proper instal-
lation (refer to scale instal-
lation instructions). Clean
the scale, if necessary.

Scale not installed
properly

Change the count direction
(pg. 8).

Wrong count
direction

Symptom

Recommended
Corrective Action

Probable Cause

Unplug the scale from the

readout and plug it into an

input for another axis.

If the scale still does not

count, the reading head

may need to be replaced. If

the scale counts on the

other axis then the display

may need to be replaced.

See LED Failure.

The scale’s reading
head is not working
properly