Troubleshooting – ACU-RITE 100S User Manual
Page 27
100S
Turning Reference Manual
TROUBLESHOOTING
24
Scale counts,
but not
correctly
Check that the scale cable
is securely connected to
the readout.
Loose cable
Check that the scale cable
is connected to the correct
input on the back of the
readout.
Wrong input
Check the scale and read-
ing head for proper instal-
lation (refer to scale instal-
lation instructions).
Scale not installed
properly
Scale won’t
count
Check the scale and read-
ing head for proper instal-
lation (refer to scale instal-
lation instructions). Clean
the scale, if necessary.
Scale not installed
properly
Change the count direction
(pg. 8).
Wrong count
direction
Symptom
Recommended
Corrective Action
Probable Cause
Unplug the scale from the
readout and plug it into an
input for another axis.
If the scale still does not
count, the reading head
may need to be replaced. If
the scale counts on the
other axis then the display
may need to be replaced.
See LED Failure.
The scale’s reading
head is not working
properly